The extended dynamic range (EDR) option for the AZURA RID 2.1L refractive index detector was investigated with a simple sample consisting of two common sugars. Calibration curves covering the range 700 μRIU to 2300 μRIU were generated with activated and deactivated EDR. A gain of about 65 % in dynamic range could be demonstrated over this range.
Overlay of an injection with 50 μL, blue=without extension, red=with extension (+100 %)
Linearity of glucose calibration with (red, R=0.9924) and without (blue, R=0.8087) EDR option
Method | HPLC |
---|
Mode | RP |
---|---|
Substances | Glucose, saccharose |
Key words | Sugar, glucose, saccharose, scale up, RID 2.1L, extended range, extended dynamic range, semi-preparative, preparative |
CAS number | 58367-01-4, 57-50-1 |