EN DE

Compact variable single wavelength UV detector with fiber optics for ASM 2.2L

Contact us
print page
Article No.: ADA07XA
Articlename: AZURA UV Detector UVD 2.1S Fiber Optics for ASM 2.2L

Compact variable single wavelength UV detector with fiber optics for ASM 2.2L


UV detector UVD 2.1S fiber optics

The AZURA UVD 2.1S is a highly competitive single variable wavelength UV detector for HPLC. It offers excellent technical specifications for routine laboratory work. With its small footprint, it is one of the smallest detectors for HPLC on the market. The UVD 2.1S comes a small housing. The installed deuterium lamp covers a wavelength range from 190 to 500 nm. The UV detector can be controlled with OpenLab EZChrom Edition®, Chromeleon®, PurityChrom®, Mobile Control® and ClarityChrom® software via LAN, as well as through analog input/output; allowing it to be integrated into almost any LC system. Due to a smart design the flow cell is easily accessible and can be changed very quickly. You can choose between a wide range of flow cells for analytical or preparative LC applications with flow rates from 10 μl/min up to 10 l/min.
This fiber optics version offers the possibility to separate the flow cell spatially from the device and thus provides enhanced security for hazardous, explosive or toxic work processes.
 

Key Features

Technical Data

Detection

Detector type Variable single wavelength UV detector
Version fiber optics version (fiber optic cables are not included)
Detection channels 1
Light source Deuterium (D²) lamp with integrated GLP chip
Wavelength range 190 - 500 nm
Spectral bandwidth 13 nm at Hα line (FWHM)
Wavelength accuracy ± 3 nm
Wavelength precision 0.7 nm (ASTM E275-93)
Noise ± 25 µAU at 254 nm (ASTM E1657-98)
Drift 400 µAU/h at 254 nm (ASTM E1657-98)
Linearity > 2.0 AU at 270 nm (ASTM E1657-98)
Maximum data rate 50 Hz (LAN)/20 Hz (Analog)/10 Hz (RS-232)
Flow cell Not included (see Accessories / Spare parts)
Time constants 0.00 / 0.02 / 0.05 / 0.1 / 0.2 / 0.5 / 1.0 / 2.0 s
Integration time Automatic
Leak sensor via ASM 2.2L

Communication

Inputs Autozero, Start (IN), Error (either IN or OUT), 0 - 10 V Analog IN
Outputs Error (either OUT or IN)
Analog inputs Wavelength 0 - 10 V via ASM 2.2L
Analog outputs 1 x ± 2.5 V scalable, 20 bit
Control Digital:LAN,RS-232, remote connector/Analog: wavelength control/Manual: front panel
Interfaces LAN, multi-pin connector, analog (RCA cinch connector) of the ASM 2.2L

Technical parameters

GLP Lamp operating hours
Ambient conditions temperature range: 4 - 40 °C, 39.2 - 104 °F air humidity: below 90%, non-condensing

General

Power supply External: input 100 - 240 V, output24 V DC, 60 W
Dimensions 112 x 132 x 260 mm (W x H x D)
Weight 2.5 kg

Other

Note Assistants with the following configuration are not permitted: 1) more than two pump modules - a high pressure gradient is not supported 2) more than one UV detector 3) without plug-in modul

Documents and more information

Certified Quality 

KNAUER is known for the quality of its products and has been certified several times. We produce high-tech laboratory instruments for research, routine analysis and production.
See all Quality certificates

In addition, we have extended documentation about all our products, including CE Declarations of Conformity.
See all CE Declarations of conformity
 

Innovation and quality - since 1962. Made in Germany.

Do you need more details or a quote?

Call us or drop us a line.

+49 30 809727-0

Mo – Fr
07:00 – 17:00 CE(S)T

sales@knauer.net

24 / 7
 

Technical data or prices are subject to change without notice. Prices may vary by country and do not include taxes, customs duties or delivery. All trademarks are the property of their respective owners.Our general terms and conditions apply: www.knauer.net/terms

Follow us on: