Our pump heads are made from materials, which are resistant to common HPLC solvents. You can use all pumps under standard conditions (1 < pH < 13, not oxidative or reductive reagent).
If you are using buffer or any high concentrated salt solutions, you should use ceramic or titanium pump heads. For liquids with a pH < 1, you can use the Hastelloy version of the 10 or 50 ml pump head.
In case your eluent is more aggressive (oxidative/reductive or a pH > 13) you have to use Kel-F / FFKM upgrade Kits:
Pump head size | Article number |
---|---|
10 ml | A5821-1 |
50 ml | A5821-2 |
100/250 ml | A58211 |
500/1000 ml | A58212 |
Get a FREE recommandation which pump fits your application best!