Designed as an extension for the Smartline Pump 1050, the Smartline Manager 5050 adds quaternary gradient functionality by means of four precisely controlled eluent proportioning valves and a high efficiency 4-channel degasser. The Manager 5050 can also be optionally equipped with an A/D-D/A interface for integration of third-party equipment into your Smartline HPLC system. The Smartline Manager 5050 (A5331) with low pressure gradient, interface and degasser module, analytical, including accessories.

Technical Data

LPG module

Number of eluents 4
Gradient range 0–100 %
Gradient progression 1 % increments
Gradient accuracy ± 0.5 %
Gradient precision < 1 % RSD
LPG maximum flow rate 50 ml/min
LPG control by Smartline Pump 1050 or through ChromGate® software; ClarityChrom® software

Degasser module

Degasser channels 4
Max. flow rate/channel up to 10 ml/min
Degassing method gas permeation through Teflon® AF amorphous fluoropolymer membrane
Degassing efficiency < 0.5 ppm dissolved O² at 1 ml/min
Degassing chamber volume 285 µl per channel
Solvent applicability universal, except fluorinated hydrocarbons
Wetted materials PEEK, Tefzel®, Teflon® AF
Vacuum chamber polypropylene and PEEK
Vacuum pump low hysteresis operation > 6.2 kPa : 400 cycles/min; < 6.2 kPa : 60 cycles/min
Chamber material polypropylene and stainless steel

Communication

A/D-D/A interface 24 bit
Analog inputs 4 x multiplexed, bipolar, differential, 24 bit resolution, maximum input range: ± 2.56 V, programmable by software (same setting for all channels), data acquisition rate: 1 Hz –10 Hz
Data acquisition rate 1–10 Hz
Analog outputs 4 x single ended unipolar, output range: 0 to +10 V, 12 bit resolution, 2.5 mV steps
Digital inputs 4 inputs, grounded contact or TTL/CMOS or TTL/Open collector compatible
Digital outputs 4 outputs, single pole double throw (SPDT) electromechanical relays

Technical parameters

Ambient conditions temperature range: 4–40 °C; 39.2–104 °F air humidity: below 90 % humidity (non condensing)

General

Power supply via Smartline Pump 1050
Dimensions 226 x 135 x 410 mm (W x H x D)
Weight max. 4.8 kg

Documents and more information

Certified Quality 

KNAUER is known for the quality of its products and has been certified several times. We produce high-tech laboratory instruments for research, routine analysis and production.
See all Quality certificates

In addition, we have extended documentation about all our products, including CE Declarations of Conformity.
See all CE Declarations of conformity
 

Innovation and quality - since 1962. Made in Germany.

Do you need more details or a quote?

We are happy to help you

+49 30 809727-0

Mo – Fr
07:00 – 17:00 CE(S)T

sales@knauer.net

24 / 7
 

Technical data or prices are subject to change without notice. Prices may vary by country and do not include taxes, customs duties or delivery. All trademarks are the property of their respective owners.Our general terms and conditions apply: www.knauer.net/terms

Follow us on: