EN DE

Biocompatible, Low pressure gradient FPLC pump

Contact us
print page
Article No.: APH69EB
Articlename: AZURA P 6.1L

Biocompatible, Low pressure gradient FPLC pump


AZURA P 6.1L Low Pressure Pump with 10 ml pump head, Ceramic, without Degasser

The AZURA® Pump P 6.1L uses technology to overcome the challenges of pumping LC solvents at high pressure and defined flow rates. This pump is designed to fulfill the needs for either high pressure or low pressure gradient forming. The pump can deliver flow in the range of 0.001–50 ml/min at pressures up to 1000 bar (depending on model). The AZURA® binary pump contains two identical high pressure pumps, a 2×2-channel inlet solvent selection valve and the low-volume AZURA® mixer. The AZURA® quaternary pump contains one high pressure pump and an integrated LPG mixing block with a 4-channel inlet solvent selection valve and a low-dwell volume AZURA® mixer. The integrated degasser and AZURA® inline filter are completing this analytical AZURA® P 6.1L HPLC pump and turn it into a work horse in the lab. This pump is also available with wetted materials made from ceramic, PEEK and titanium for biocompatible applications.

Technical Data

General information

Max. flow rate 10 ml/min
Flow rate range 0.001–10 ml/min
Flow rate increment 0.001 ml/min
Pressure range
Maximum delivery pressure [psi] 5800 psi
Maximum delivery pressure [MPa] 40 MPa
Maximum delivery pressure [bar] 400 bar
Pump head materials Ceramic
Gradient Low pressure gradient (quaternary)
Leak management Yes
Wetted materials sapphire, ruby, ceramic, FKM
Pump features Active pulsation compensation, RFID for automatic pump head detection, Integrated pressure sensor, Purge valve (manual), Leak management, Piston seal wash

Detailed information

Best working conditions 0.1–8.0 ml/min
Continuous working conditions 0.1–4.0 ml/min
Flow rate accuracy ± 0.25 %
Flow rate accuracy conditions measured at 5–80 % of flow range, using ethanol
Flow rate precision ≤ 0.04 % RSD or 0.008 min SD (whichever is greater)
Flow rate precision conditions based on the retention time at constant room temperature
Gradient range 0–100 %
Gradient range (recommended) 1–99 %
Gradient accuracy ± 0.3 % at 1 ml/min, 150 bar ethanol/caffeine tracer) ± 2 % (1–99 %, measured at 5–50 % of the flow range, water/caffeine tracer)
Gradient precision < 0.1% RSD at 1 ml/min, 0.5 % RSD overall, based on retention time at constant room temperature
Gradient minimum increment 0.1 %
Mixing volume 250 µl
Delay volume 350 µl
Degasser No
Pulsation < 2 % Amplitude (typically: < 1.3 %) or < 3 bar (0.3 MPa), whatever is greater, at 1 ml/min ethanol, at all pressures > 10 bar (1 MPa, 147 psi)
Pulsation compensation Active pulsation compensation
Active piston seal wash Yes
Piston seal washing Yes
System protection Soft start, Pmin und Pmax are programmable
Pump Outlet thread Female UNF 10-32 (coned)
Pump head inlet (standard) UNF 1/4-28 Thread
Pump head outlet (standard) UNF 10-32 Thread

Communication

Interfaces LAN, Pin header connectors (Analog IN, Start In, Error IN)
Control LAN, Analog and event control, Mobile Control
Analog inputs Flow rate, 0–10 V via pin header connectors
Analog outputs 8 event outputs (TTL, OC, Relais) and 24 V

General

Power supply 100–240 V; 50–60 Hz; Maximum power consumption 100 Watt
Dimensions 361 mm × 208.2 mm × 523 mm (W × H × D)
Leak sensor Yes
Ambient conditions 4–40 °C (39.2–104 °F) Air humidity below 90 %, non-condensing

Documents and more information

Certified Quality 

KNAUER is known for the quality of its products and has been certified several times. We produce high-tech laboratory instruments for research, routine analysis and production.
See all Quality certificates

In addition, we have extended documentation about all our products, including CE Declarations of Conformity.
See all CE Declarations of conformity
 

Innovation and quality - since 1962. Made in Germany.

Do you need more details or a quote?

We are happy to help you

+49 30 809727-0

Mo – Fr
07:00 – 17:00 CE(S)T

sales@knauer.net

24 / 7
 

Technical data or prices are subject to change without notice. Prices may vary by country and do not include taxes, customs duties or delivery. All trademarks are the property of their respective owners.Our general terms and conditions apply: www.knauer.net/terms

Follow us on: