EN DE

How does the eluting strength of my solvent influence the separation?

The eluting strength is classified with the eluotropic series. In this series the solvents are arranged in order of their relative abilities to effect elution or respectively the order of solvents runs in the order of their polarity. Low polarity n-hexane ⇒toluene ⇒chloroform ⇒dichloromethane ⇒acetone ⇒dimethyl sulphoxide ⇒acetonitrile ⇒isopropanol/n-propanol ⇒ ethanol ⇒methanol ⇒water ⇒salts and buffers High polarity If you have a mixture of acetonitrile/water 50:50 (v/v) and want to switch to methanol, you need about 20% more methanol to get the same eluting strength. You also have to consider the possible interaction between the different molecules or functional groups of your sample, mobile phase and stationary phase.

Wir sind persönlich für Sie da –
in ganz Deutschland, Österreich und der Schweiz!

Haben Sie Fragen zu unseren Produkten oder benötigen Sie ein Angebot?
Kontaktieren Sie uns – wir helfen Ihnen gerne weiter!

Braunschweig

Dr. Diana Rohrbeck
 

Tel: +49 151 15943077
Meeting vereinbaren

E-Mail schreiben

Berlin

Oskar Klak, M.Sc.
 

Tel: +49 170 2862155
Meeting vereinbaren

E-Mail schreiben

Frechen

Sandra Lehmann, M.Sc.
 

Tel: +49 171 4113554
Meeting vereinbaren

E-Mail schreiben

Leipzig

Dipl.-Ing. (FH) Micha Hinderer
 

Tel: +49 151 43536226
Meeting vereinbaren

E-Mail schreiben

Frankfurt am Main

Can Neis, BE
 

Tel: +49 171 1227074
Meeting vereinbaren

E-Mail schreiben

München

Dipl.-Ing. (FH) Christian Ulbricht
 

Tel: +49 170 9275967
Meeting vereinbaren

E-Mail schreiben

Österreich & Schweiz

Marius Segl, M.Sc.
 

Tel: +43 676 3806550
Meeting vereinbaren

E-Mail schreiben

Finden Sie Ihre/n Ansprechpartner/in
in Deutschland ...

... nach Postleitzahl

Brauchen Sie mehr Informationen?

Kontaktieren Sie uns, wir helfen Ihnen gerne!

+49 30 809727-0

Mo – Fr
07:00 – 17:00 CE(S)T

sales@knauer.net

24 / 7
 

Besuchen Sie uns auf: