EN DE

What is the bandwidth of a diode array detector?

When a specific wavelength is set on an diode array detector (DAD), the total number of wavelengths actually registered by the photodiode is refered to as the bandwidth. For example, a wavelength set at 254 nm with a bandwidth of 4 nm (254/4 nm) results in average absorption of 252-256 nm. The selection of the bandwidth is a balance between sensitivity and selectivity. Narrower bandwidths increase selectivity, whereas broad bandwidths increase sensitivity. The default bandwidth for the AZURA DAD family is set to 8 nm.The bandwidth should not be confused with the spectral bandwidth, which is a physical property of a given detector.

Wir sind persönlich für Sie da –
in ganz Deutschland, Österreich und der Schweiz!

Haben Sie Fragen zu unseren Produkten oder benötigen Sie ein Angebot?
Kontaktieren Sie uns – wir helfen Ihnen gerne weiter!

Braunschweig

Dr. Diana Rohrbeck
 

Tel: +49 151 15943077
Meeting vereinbaren

E-Mail schreiben

Berlin

Oskar Klak, M.Sc.
 

Tel: +49 170 2862155
Meeting vereinbaren

E-Mail schreiben

Frechen

Sandra Lehmann, M.Sc.
 

Tel: +49 171 4113554
Meeting vereinbaren

E-Mail schreiben

Leipzig

Dipl.-Ing. (FH) Micha Hinderer
 

Tel: +49 151 43536226
Meeting vereinbaren

E-Mail schreiben

Frankfurt am Main

Can Neis, BE
 

Tel: +49 171 1227074
Meeting vereinbaren

E-Mail schreiben

München

Dipl.-Ing. (FH) Christian Ulbricht
 

Tel: +49 170 9275967
Meeting vereinbaren

E-Mail schreiben

Österreich & Schweiz

Marius Segl, M.Sc.
 

Tel: +43 676 3806550
Meeting vereinbaren

E-Mail schreiben

Finden Sie Ihre/n Ansprechpartner/in
in Deutschland ...

... nach Postleitzahl

Brauchen Sie mehr Informationen?

Kontaktieren Sie uns, wir helfen Ihnen gerne!

+49 30 809727-0

Mo – Fr
07:00 – 17:00 CE(S)T

sales@knauer.net

24 / 7
 

Besuchen Sie uns auf: