EN DE

Size exclusion chromatography of polylactic acid in three different solvents

Applikationsnr.: VEV0085

Size exclusion chromatography of polylactic acid in three different solvents

Several different polylactic acid (PLA) samples have been successfully examined with respect to their molar mass and polydispersity (PD) using a KNAUER HPLC system in combination with a gel permeation chromatography (GPC) column. Calibrations of polystyrene (PS) and polymethyl methacrylate (PMMA) in three different solvents, tetrahydrofuran (THF), chloroform and ethyl acetate (EtOAc) have been recorded. In this study a 20 kDa PLA standard and two real samples were evaluated using these calibrations and the results compared with one another. It could be shown that EtOAc is suitable as an environmentally friendly alternative for the analysis of low molecular weight P(D,L)LA samples.

Technische Daten

Spezifikationen

Methode HPLC
Modus GPC
Substanzen polylactic acid
Schlüsselwörter PLA, polylactic acid, GPC, SEC, polymer, biopolymers, bioplastic, biodegradable, universal calibration, PS, polystyrene, PMMA, polymethylmethacrylate, green SEC, green GPC, biobased plastic
Version Version 1/ 11/2021

Dokumente und weitere Informationen

Benötigen Sie Hilfe bei Ihrer Applikation?

Wir helfen Ihnen gerne!

Besuchen Sie uns auf: