How can I prevent peak broadening in an FPLC system?

Peak broadening due to higher flow rates in the middle of tubing

Peak broadening is the effect that a product is not staying concentrated in its injection volume, but is diffusing while passing through the bio purification system. This effect can impair the resolution of your separation. The reason behind peak broadening is that the flow rate in the middle of the tubing is higher than at the tubing walls, which is visible for low concentrated peaks most. One can avoid peak broadening by shortening the tubing length of the overall system and reducing inner tubing diameter. But keep in mind, that smaller inner tubing diameters are also increasing the system pressure.

Wir sind persönlich für Sie da –
in ganz Deutschland, Österreich und der Schweiz!

Haben Sie Fragen zu unseren Produkten oder benötigen Sie ein Angebot?
Kontaktieren Sie uns – wir helfen Ihnen gerne weiter!

Braunschweig

Dr. Diana Rohrbeck
 

Tel: +49 151 15943077
Meeting vereinbaren

E-Mail schreiben

Berlin

Oskar Klak, M.Sc.
 

Tel: +49 170 2862155
Meeting vereinbaren

E-Mail schreiben

Frechen

Sandra Lehmann, M.Sc.
 

Tel: +49 171 4113554
Meeting vereinbaren

E-Mail schreiben

Leipzig

Dipl.-Ing. (FH) Micha Hinderer
 

Tel: +49 151 43536226
Meeting vereinbaren

E-Mail schreiben

Frankfurt am Main

Can Neis, BE
 

Tel: +49 171 1227074
Meeting vereinbaren

E-Mail schreiben

München

Dipl.-Ing. (FH) Christian Ulbricht
 

Tel: +49 170 9275967
Meeting vereinbaren

E-Mail schreiben

Österreich & Schweiz

Marius Segl, M.Sc.
 

Tel: +43 676 3806550
Meeting vereinbaren

E-Mail schreiben

Finden Sie Ihre/n Ansprechpartner/in
in Deutschland ...

... nach Postleitzahl

Brauchen Sie mehr Informationen?

Kontaktieren Sie uns, wir helfen Ihnen gerne!

+49 30 809727-0

Mo – Fr
07:00 – 17:00 CE(S)T

sales@knauer.net

24 / 7
 

Besuchen Sie uns auf: