EN DE

Analysis of Poly [(R)-3-hydroxybutyric acid] in chloroform using GPC and universal calibration

Applikationsnr.: VEV0084

Analysis of Poly [(R)-3-hydroxybutyric acid] in chloroform using GPC and universal calibration

Two Poly [(R)-3-hydroxybutyric acid] (PHB) standards (10 - and 500 kDa) were analysed in chloroform with SEC using universal calibration. For this purpose, a KNAUER system was used in combination with a linear GPC column. For the determination of the number average molecular mass (Mn) and the mass average molecular mass (Mw) as well as the poly dispersity (PD), polystyrene (PS) and polymethylmethacrylate (PMMA) calibrations were performed in chloroform. It could be shown that this system can be successfully used for the analysis of the low molecular weight standard. The molar mass of the 500 kDa standard, on the other hand, was greatly underestimated with this method, but this is probably related to the poor solubility or the sample preparation of the higher molecular weight standard.

Technische Daten

Spezifikationen

Methode HPLC
Modus GPC
Substanzen Poly [(R)-3-hydroxybutyric acid]
Schlüsselwörter PHB, Poly [(R)-3-hydroxybutyric acid], GPC, SEC, polymer, biopolymers, bioplastic, biodegradable, universal calibration, PS, PMMA, green SEC, green GPC, biobased plastic, polydispersity
CAS Nummer 26063-00-3
Version Version 1/ 10/2021

Dokumente und weitere Informationen

Benötigen Sie Hilfe bei Ihrer Applikation?

Wir helfen Ihnen gerne!

Besuchen Sie uns auf: