Science with Passion
Application No.: VTN0040 Version 1 02/2025
Benefits of KNAUER´s trouble-free analytical capillaries in terms of HPLC and UHPLC system performance
M. Myrach, J. Kramer; m.myrach@knauer.net
KNAUER Wissenschaftliche Geräte GmbH, Hegauer Weg 38, 14163 Berlin
Photo: KNAUER
Summary
KNAUER has introduced new capillaries, fittings and Start-Up Kits for all AZURA® Analytical HPLC and UHPLC/ULDC systems. The goal was to provide a “trouble-free” solution that is easy to handle and reusable many times. But does it affect the performance of the system?
Introduction
Every (U)HPLC user needs to install and replace capillaries and fittings on a regular basis. These comparably small and inexpensive supplies can have a major influence on both the user experience and the system performance. Both factors are important regarding the outcome of every single analysis.
KNAUER decided to use the well-established IDEX MarvelXACTTM capillaries and fittings for all AZURA® Analytical systems. They are finger tight up to UHPLC pressure, so that no tools are required. They auto adjust to various port depths and do not depend on ferrules. The sealing at the bottom of the port is achieved without complex techniques, which significantly reduces required torque (Fig.1).
The trouble-free connection system eliminates the risk of under- and over-tightening, with a patented torque-limiting mechanism. This unique feature provides a haptic “click” feedback when it reaches the optimum torque, assuring a perfect connection for every user, independent from the expertise level. These capillaries can be connected and disconnected more than 100 times.
Fig. 1 Trouble-free connection system, sealing at the front side
Start-up kits are available for any standard AZURA® Analytical system. These kits have been designed for most convenient installation with a comparable back pressure rating to the former used kits, providing equal or better system performance at the same time. The benefits of user experience are clear, but what about system performance compared to the conventional capillary/ferrule connection systems?
In this tech note, we show the results and unravel the mystery of whether this design worked as planned in laboratory reality.
System build-up
Conventional K-Connect, which uses the classical system of ferrule and fitting, and trouble-free MarvelXACTTM connections were tested on an AZURA® Analytical HPLC 862 bar system and a UHPLC 1240 bar system. The system build-up (Fig. 2), HPLC method, solvents and sample were kept the same, only the capillaries and fittings were exchanged according to the instructions of the start-up kits. A well-established method used for system performance verification was used and adapted for these tests. After an isocratic run to determine the system performance based on the changed extra-column volume (ECV), also a gradient run was performed to see the impact of the changed dwell volume (see Tab.1 and Tab.2). The extra column volume is defined as the volume between the point of the injection and the point of detection. It relates to the band spread, and every connection, tubing, or fitting in the post-injection flow path contributes to the ECV. The dwell volume, also known as gradient delay volume, is determined by the pump characteristics. It is defined as the volume from the point of mobile phase mixing to the inlet of the column.
Fig. 2 AZURA® Analytical system build-up with installed capillaries 1, 2 and 3
Tab. 1 Capillaries for AZURA® Analytical HPLC 862 bar systems
Tab. 2 Capillaries for AZURA® Analytical UHPLC 1240 bar systems
Results
For the AZURA® Analytical HPLC 862 bar system, the overall pressure including the column rose from 60 to 70 bar when changing to the trouble-free capillary system. The isocratic run to determine the system performance gave similar results for both capillary systems (Fig.3). Retention times as well as theoretical plates and peak shapes were comparable (Tab.3).
Tab. 3 Results for the AZURA® Analytical HPLC 862 bar system
Fig. 3 Chromatograms of isocratic run on AZURA® Analytical HPLC 862 bar system; blue: K-Connect; green: MarvelXACTTM; (1) Butyl benzoate; (2) Pentyl benzoate
Fig. 4 Zoom in overlay chromatograms of isocratic run on AZURA® Analytical HPLC 862 bar system; blue: K-Connect; green: MarvelXACTTM; (1) Butyl benzoate; (2) Pentyl benzoate
The small variation in retention time (Fig.4) can be explained by the lower extra column volume for the new capillary kits. As it is an isocratic run, the slightly higher dwell volume does not play a role here. The lower extra column volume should also result in a better performance and therewith a higher number of theoretical plates. The used HPLC column was chosen in standard HPLC dimensions of 150 x 4 mm with a resulting column void volume of around 1.28 ml. The extra column volume therewith only plays a minor role. The expected effect is more relevant in UHPLC where smaller columns are applied, and the extra column volume becomes more significant compared to the column void volume. As expected, there are no major differences in peak retention using gradient elution because the dwell volume does not differ significantly for both capillary kits (Fig.5 and Fig.6).
Fig. 5 Chromatograms of gradient run on AZURA® Analytical HPLC 862 bar system; blue: K-Connect; green: MarvelXACTTM; (1) Butyl benzoate; (2) Pentyl benzoate
Fig. 6 Zoom in overlay chromatograms of gradient run on AZURA® Analytical HPLC 862 bar system; blue: K-Connect; green: MarvelXACTTM; (1) Butyl benzoate; (2) Pentyl benzoate
For the AZURA® Analytical UHPLC 1240 bar system, the overall pressure including the column did not increase when changing the capillaries. The system performance was enhanced using the new capillary system. Retention times and peak shapes were comparable, while the number of theoretical plates was improved significantly with the new capillary system (Fig.7 and Fig.8).
Fig. 7 Chromatograms of isocratic run on AZURA® Analytical UHPLC 1240 bar system; blue: K-Connect; green: MarvelXACTTM; (1) Butyl benzoate; (2) Pentyl benzoate
Fig. 8 Zoom in Chromatograms of isocratic run on AZURA® Analytical UHPLC 1240 bar system; blue: K-Connect; green: MarvelXACTTM; (1) Butyl benzoate; (2) Pentyl benzoate
In UHPLC, the extra column volume plays a major role as column void volume becomes smaller. A common UHPLC column in the dimension 50 x 2 mm ID was chosen with a column void volume of about 0.11 ml. By reducing the extra column volume and especially by cutting the volume between column outlet and detector in half, a significant rise in theoretical plates was observed (see Tab. 4).
Tab. 4 Results for the AZURA® Analytical UHPLC 1240 bar system
For the gradient run, the results were similar for retention times of the two peaks (Fig. 9 and Fig. 10). This was expected, because the dwell volume does not differ significantly for both capillary kits.
Fig. 9 Chromatograms of gradient run on AZURA® Analytical UHPLC 1240 bar system; blue: K-Connect; green: MarvelXACTTM; (1) Injection peak; (2) Butyl benzoate; (3) Pentyl benzoate
Fig. 10 Zoom in Chromatograms of gradient run on AZURA® Analytical UHPLC 1240 bar system; blue: K-Connect; green: MarvelXACTTM; (2) Butyl benzoate; (3) Pentyl benzoate
In the gradient run, a little shift of retention times can be seen. With the new capillary system, both peaks elute a little later than with the former K-Connect system. In Fig. 9, also an injection peak can be seen. The injection peak is detected a bit earlier with the new capillaries compared to the former system. Tab. 2 gives the explanation for these phenomena: the retention time of the injection peak is not affected by the eluent composition but as the extra column volume is smaller with the new capillary kit, the injection peak is detected earlier compared to the K-connect capillary kit. At the same time, the analytes are affected by eluent composition. This is one of the reasons why they can be separated on this stationary phase. Because the dwell volume is slightly higher with the new capillary kit, the higher eluting eluent composition arrives at the column later compared to the former capillary kit. If a smaller dwell volume and therewith a faster gradient formation is needed, capillaries with smaller volumes can also be applied and must be added to the standard start-up kit. Always be aware that they will have a smaller inner diameter which will result in an increase in the overall system backpressure.
Conclusion
Applying KNAUER´s trouble-free start-up kits for AZURA® Analytical HPLC and UHPLC/ULDC systems brings benefits to everyday laboratory life. The ease of use and the reusability of capillaries and fittings as well as the resulting system performance contribute to a positive user experience. These kits can replace the former recommended K-Connect kits without any limitations. Work tool-free and trouble-free while keeping the known HPLC or UHPLC system performance or even enhancing it.
Material and Methods
Tab. 5 System configuration AZURA® Analytical HPLC 862 bar
Instrument | Description | Article No. |
Pump | AZURA P 6.1 L HPG, 10 ml SST pump head | |
Autosampler | AZURA AS 6.1L, Standard, 862 bar | |
Column Thermostat | AZURA CT 2.1 | |
Column | Eurospher II 100-5 C18, 150 x 4 mm ID with integrated precolumn | |
Detector | AZURA DAD 2.1L | |
Flow cell | Analytical PressureProof Flow cell cartridge, 10 mm path length, 10 µl volume | |
Capillaries K-Connect | Start up Kit including: Pump – Autosampler: Length 700 mm, | - - - - |
Capillaries MarvelXACTTM | Start up Kit including: Pump – Autosampler: Length 600 mm, | |
Software | ClarityChrom 9.0.0 – Workstation, autosampler control included | |
Software | ClarityChrom 9.0.0 – PDA extension |
Tab. 6 System configuration AZURA® Analytical UHPLC 1240 bar
Instrument | Description | Article No. |
Pump | AZURA P 8.1L, HPG, 5 ml SST pump head | |
Autosampler | AZURA AS 6.1L, standard, 1240 bar | |
Column Thermostat | AZURA CT 2.1 | |
Column | Zorbax RRHD 1,8 µm SB-C18 50 x 2 mm | - |
Detector | AZURA DAD 6.1L | |
Flow cell | Analytical PressureProof Flow cell cartridge, 10 mm path length, 10 µl volume | |
Capillaries K-Connect | Start up Kit including: Pump – Autosampler: Length 700 mm, | - - - - |
Capillaries MarvelXACTTM | Start up Kit including: Pump – Autosampler: Length 600 mm, | |
Software | ClarityChrom 9.0.0 – Workstation, autosampler control included | |
Software | ClarityChrom 9.0.0 – PDA extension |
Tab. 7 Isocratic method for AZURA® Analytical HPLC 862 bar
Tab. 8 Gradient method for AZURA® Analytical HPLC 862 bar
Tab. 9 Isocratic method for AZURA® Analytical UHPLC 1240 bar
Tab. 10 Gradient method for AZURA® Analytical UHPLC 1240 bar
Application details
Method | HPLC, UHPLC |
Mode | RP |
Substances | Butyl benzoate, Pentyl benzoate |
CAS number | 136-60-7, 2049-96-9 |
Version | Application No.: VTN0040 | Version 1 02/2025 | ©KNAUER Wissenschaftliche Geräte GmbH |